STEP #1 - The cassette must be opened and cooled before removing the cassette seal.
See Figure A.
STEP #2 - Start with any of the (4) corners and begin prying the cassette seal loose using the RPI Cassette Seal Removal
Tool (RPI Part #RPT372) and by pulling down and out at the same time.
See Figures B and C.
STEP #3 - Once the first corner is free, take hold of the cassette seal and pull down and out in order to remove it from the
cassette groove. Be careful not to tear the (2) locating tabs in each corner of the cassette seal.
See Figures B
and C.
STEP #4 - Once the cassette seal has been removed, it is important to inspect, clean and remove any residue and/or debris
from the cassette groove and outer edge. Check for dings and dents on both halves of the cassette; if present, a
proper seal may not occur.
See Figure D.
STEP #5 - Using (1) package of RPI Liquid Soap (RPI Part #RPS287), lube the cassette seal completely.
Note: It is not nec-
essary to lube the inside of the cassette seal. See Figure E.
STEP #6 - Now you are ready to install the cassette seal.
STEP #7 - When installing the cassette seal, it is best to start in the back corner, nearest the port holes of the cassette.
Slip the cassette seal under and into the corner, placing the locating tabs at the edge of the open corner. Now
align the port holes and their locating tabs with the cassette. The locating tabs must protrude through the
square notches of the cassette lid.
See Figure F.
STEP #8 - Install the other corners in the same manner. Now insert the cassette seal into the groove of the cassette – one
side at a time.
See Figures G and H.
STEP #9 - As you move along installing the cassette seal, check the corners to ensure that the locating tabs stay in place.
Note: The sides should be inserted until smooth – without waves or bumps. See Figure I.
STEP #10 - After you have completely installed the cassette seal, make a final inspection of all corners and all locating
tabs. Remove excess soap from the port holes as the soap can cause a build up on the steam ports and in
turn cause a bad seal.
STEP #11 - Run a few test cycles to check for steam leaks, then apply Sci-Dry™ (RPI Part #SCA054).
Note: During the
test cycles, it is normal to hear a hissing sound at first. The sound is from the excess liquid soap expelling
and the cassette seal seating in place.
STEP #12 - After running the test cycles, if steam still leaks, reinstall the cassette seal and run additional test cycles. If
steam leaks still occur, it is possible that the problem was not a faulty cassette seal, but something else, so
refer to the operator’s manual or contact your service professional.
FIGURE A FIGURE B
FIGURE D FIGURE E
FIGURE G FIGURE H FIGURE I
Locating
Tabs
Port
Holes
Residue
and/or
Debris
FIGURE C
FIGURE F
USAGE TIPS
✔
✔
✔
✔
✔
✔
✔
✔
✔
✔
✔
✔
Always note the position of the cassette when inserting
it into the unit – if the cassette is installed upside down,
the thermocouple will bend and the sterilizer will become
inoperative.
The OEM recommends lubrication of Cassette Seal every
30 days and replacement every 6 months or 500 cycles
– whichever comes first. Clean the outer edge of cas-
sette tray and the exposed area of the Cassette Seal with
RPI Liquid Soap (RPI Part #RPS287) weekly. Apply Sci-
Dry™ (RPI Part #SCA054) every 10 cycles and after
cleaning.
Always keep the inside of the cassette clean. Remove
residue and/or debris from the metal edge of the lower
half of the cassette on a regular basis.
4.0 - INSTALLATION AND MONTHLY MAINTENANCE FOR CASSETTE SEAL
Cassette Seal (RPI Part #SCS001) to fit the SciCan STATIM 2000
Cassette Seal (RPI Part #SCS029) to fit the SciCan STATIM 5000
Sci-Dry™ (RPI Part #SCA054) should be applied after
every 10 cycles and after cassette cleaning. Coat all in-
ternal surfaces of cassette. Sci-Dry™ minimizes spotting
and allows instruments to dry more efficiently.
Watch for steam leaks from the cassette bay as they can
cause damage to the microswitch and other internal com-
ponents, thus rendering the sterilizer inoperative.
Refill the reservoir with distilled water only. At the same
time that the reservoir is refilled, the condenser waste
bottle should be emptied and refilled. Refill the condenser
waste bottle with tap water to the Minimum Fill Level
as
shown on the bottle.
(Note: Never operate sterilizer when
water level is above the Maximum Fill Line or below the
Minimum Fill Line.)