6.0 – LCD ERROR CODES
(continued)
STAGE #3 – PRESSURIZING
The pressure increases as the cassette temperature rises to the factory set temperatures for the “Unwrapped”, “Wrapped” and
“Rubber & Plastic” cycles. The vent solenoid will open and close several times to maintain proper cassette steam temperature,
expelling steam/air into the condenser waste bottle.
LCD MESSAGE
(For Model 2000 only)
CHECK CASSETTE #3 (CYCLE FAULT #3)
The cassette fails to pressurize or reach temperature
(~95˚ - 110˚C) within ~70 seconds (or within the time out
period).
THINGS TO CHECK
Clean and lubricate the cassette seal – see page 10,
2.7 - Liquid Soap (RPI P
art #RPS287). Replace the
Cassette Seal if necessary (RPI Part #SCS001 fits model
2000), and (RPI Part #SCS029 fits model 5000).
• Inspect and clean steam ports on probe bracket.
• Solenoid valve plunger may be compromised by debris
or may be worn. Check the plunger inside the flute/bon-
net for proper movement. Replace Solenoid Plunger Kit
(RPI Part #SCK003) if necessary. See page 41, Figure
9.4 - Solenoid Valve.
LCD MESSAGE
CHECK CASSETTE #4 (CYCLE FAULT #4)
The "Sterilization Stage" fails to begin within specified
time noted in the chart below.
THINGS TO CHECK
Refer to Stage #3 - Pressurizing: "Check Cassette #3
(Cycle Fault #3)".
Check Safety Valve (70 PSI) (RPI Part #SCV004) for leaks
and replace if necessary.
Leaky check valve – see page 24, Service Tip #2. In-
spect the compressor air filter. If it is wet, replace the
Check Valve Kit (RPI Part #SCK011), the Air Filter (RPI Part
#SCF002) and/or the Biological Filter (RPI Part #SCF032).
LCD MESSAGE
(For Model 2000 only)
CHECK CASSETTE #5
The steam generator/boiler may not be getting enough
water or steam generator/boiler is suspect.
THINGS TO CHECK
A weak pump due to dirty mesh filters in the pump
fittings or dirty Water Filter (RPI Part #SCF034). Check
pump’s operation by using the Pump Tester Bottle Kit
(RPI Part #SCK024) – see pages 6 and 7, 2.3 - Pump
Tester Bottle (RPI Part #SCK024). If pump passes test,
boiler must be replaced.
• Clean or calibrate steam generator/boiler see page 39
Figure 9.2 or page 40 Figure 9.3.
STATIM Model # Should begin "Sterilization Stage"
2000 101102 Within ~ 10 minutes of reaching
S/N "A" temperature (
~ 95˚-110˚C
)
2000 121101 Within ~ 6 minutes of reaching
temperature (
~ 95˚-110˚C
)
5000 01-201103 Within ~ 6 minutes of reaching
temperature (
~ 95˚-110˚C
)
6.0 – LCD ERROR CODES
(continued)
STAGE #4 – STERILIZATION
The temperature and pressure is maintained for the chosen cycle. Air/steam is being purged from the cassette as the vent
solenoid valve opens and closes several times to maintain optimum temperature. The pump continues to pump distilled water
into the steam generator/boiler which in turn pumps saturated steam into the cassette. The condenser waste bottle collects
and allows the steam to cool down safely.
LCD MESSAGE
CHECK CASSETTE #6
Applies to the STATIM 2000, Model 101102 only (with
serial number that end with an "A" – without a water
quality sensor).
The steam generator/boiler may be contaminated. The
steam generator/boiler's temperature should be at least
5˚C higher than the cassette's temperature during the
Sterilization Stage.
THINGS TO CHECK
For earlier models only: The steam generator/boiler may
be contaminated by mineral deposits, calcium, etc., that
have accumulated over time. These deposits are caused
by using low quality water.
Drain the reservoir. Clean the steam generator/boiler using
Jelmar CLR
®
Cleaner (removes mineral deposits, calcium,
etc). Refill the reservoir using high quality distilled water.
See page 14, 2.11 - Water Quality Tester and page 39,
Figure 9.2 - Cleaning the Steam Generator/Boiler.
For later models only: Calibrate the steam generator/boiler
see page 40, Figure 9.3 - Steam Generator/Boiler
Calibration Procedure.
Important Note: The later STATIM models have a built-in
water quality sensor. Do not try to clean the later STATIMs'
steam generator/boiler. If you use a lesser quality water,
the LCD on the later STATIM will display "Water Quality is
not Acceptable" or "Distilled H20 only". The sensor de-
tects unacceptable levels of dissolved solids in the reser-
voir water. See page 14, 2.11 - Water Quality Tester.
LCD MESSAGE
CYCLE FAULT #6
Applies to STATIM 2000, model 121101, and STATIM
5000, model 01-201103
This “Cycle Fault #6” will appear only with a water qual-
ity sensor installed.
Steam generator/boiler calibration is needed – see page
40, Figure 9.3 - Steam Generator/Boiler Calibration
Procedure. If problem (Error code) repeats, replace steam
generator/boiler.
THINGS TO CHECK
Water level in the waste bottle may be full, or the ex-
haust path may have a restriction or blockage.
Check exhaust tubing for kinks or pinch points. Note: RPI
offers an Exhaust Tubing Kit (RPI Part #SCK017) or Ex-
haust Tubing sold by the foot (RPI Part #RPT380).
• Check the cassette tray venturi plate for blockage - see
page 42, Figure 9.5.
• Check for steam leaks.
• Solenoid valve plunger may be compromised by debris
or may be worn. Check the plunger inside the flute/bon-
net for proper movement. Replace Solenoid-Plunger Kit
(RPI Part #SCK003) if necessary see page 41, Figure
9.4 - Solenoid Valve.